PL EN
RESEARCH PAPER
Calcification of Hydrophilic Acrylic Intraocular Lenses with a Hydrophobic Surface Following Uneventful Descemet’s Stripping Automated Endothelial Keratoplasty
 
More details
Hide details
1
LASER Eye Microsurgery Centre of Warsaw, Poland
 
2
SPKSO Ophthalmic University Hospital of Warsaw, Poland
 
3
Analytical Laboratory, Faculty of Chemistry, Maria Curie-Sklodowska University of Lublin, Poland
 
4
Department of Medicinal Chemistry, Medical University of Lublin, Poland
 
5
Department of Ophthalmology, Faculty of Medicine, Medical University of Warsaw, Poland
 
 
Submission date: 2024-05-07
 
 
Acceptance date: 2024-09-15
 
 
Publication date: 2025-01-04
 
 
Ophthalmology 2024;27(4):11-20
 
KEYWORDS
ABSTRACT
Purpose:: This is a retrospective, observational case series of 43 patients who underwent Descemet’s striping automated endothelial keratoplasty with implants of hydrophilic acrylic intraocular lenses with hydrophobic surface (Lentis LS-302-Y or Lentis L-302-1). Methods:: Patients diagnosed with intraocular lens opacification after Descemet’s striping automated endothelial keratoplasty were identified from clinic records, with minimal 18-month postoperative follow-up guaranteed in 36 eyes. Analysis included demographic data, indication for Descemet’s striping automated endothelial keratoplasty, ocular comorbidities, intraocular lens specification, complications, postoperative course, incidence of lens exchange due to intraocular lens opacification, and corrected distance visual acuity (Snellen) before surgery and before and after intraocular lens opacification. Two explanted intraocular lenses were subjected to detailed light microscopy, optical profilometry, and scanning electron microscopy with energy dispersive X-ray spectroscopy (SEM/ EDS). Results:: Opacification occurred in 81% of Lentis LS-302-Y (25/31) and in 92% of Lentis L-302-1 (11/12) intraocular lenses. The morphology of surface irregularity indicates that formation begins inside the intraocular lens material, with deformation of the polymer surface is secondary to the growth of crystal granules in the anterior subsurface area. SEM/ EDS analysis revealed significant peaks for calcium and phosphate. The presence of silicon in the examined areas of an opacified intraocular lens is worth noting. The high incidence of this complication with this particular type of intraocular lens after Descemet’s striping automated endothelial keratoplasty suggests that material-related factors may play an important role in the development of calcification. Conclusion:: Because of the increased risk of opacification after Descemet’s striping automated endothelial keratoplasty, hydrophilic and hydrophilic-hydrophobic acrylic intraocular lenses should be avoided in patients with endothelial cell disorders.
REFERENCES (35)
1.
Werner L: Causes of intraocular lens opacification or discoloration. J Cataract Refract Surg. 2007; 33: 713–726.
 
2.
Werner L: Calcification of hydrophilic acrylic intraocular lenses. Am J Ophthalmol. 2008; 146930: 341–343.
 
3.
Neuhann IM, Kleinmann G, Apple DJ: A new classification of calcification of intraocular lenses. Ophthalmology. 2008; 115: 73–79.
 
4.
Yu AKF, Kwan KYW, Chan DHY, et al.: Clinical features of 46 eyes with calcified hydrogel intraocular lenses. J Cataract Refract Surg. 2001; 27: 1596–1606.
 
5.
Pandey SK, Werner L, Apple DJ, et al.: Hydrophilic acrylic intraocular lens optic and haptics opacification a diabetic patient; bilateral case report and clinicopathologic correlation. Ophthalmology. 2002; 109: 2042–2051.
 
6.
Lee DH, Seo Y, Joo CK: Progressive opacification of hydrophilic acrylic intraocular lenses in diabetic patients. J Cataract Refract Surg. 2002; 28: 1271–1275.
 
7.
Purbrick RM, Stavrakas P, Porooshani H, et al.: Calcification of a Rayner Centerflex 570H hydrophilic acrylic intraocular lens following vitrectomy for retinal detachment: a clinicopathologic report. Eur J Ophthalmol. 2010; 20 (6 November–December): 1082–1085.
 
8.
Stringham J, Werner L, Monson B, et al.: Calcification of Different Designs of Silicone Intraocular Lenses in Eyes with Asteroid Hyalosis. Ophthalmology. 2010; 117: 1486–1492.
 
9.
Jensen MK, Crandall AS, Mamalis N, et al.: Crystalization on intraocular lens surface associated with the use of Healon GV. Arch Ophthalmol. 1994; 112: 1037–1042.
 
10.
Sher JH, Gooi P, Dubinski W, et al.: Comparison of the incidence of opacification of hydroview hydrogel intraocular lenses with the ophthalmic viscosurgical device used during surgery. J Cataract Refract Surg. 2008; 34: 459–464.
 
11.
Yu AKF, Ng ASY: Complications and clinical outcomes of intraocular lens exchange in patients with calcified hydrogel lenses. J Cataract Refract Surg. 2002; 28: 1217–1222.
 
12.
Fung SS, Sykakis E, Islam NM, et al.: Intraocular Lens Opacification following Intracamerai Injection of Recombinant Tissue Plasminogen Activator to Treat Inflammatory Membranes after Cataract Surgery. J Ophthalmol. 2015; 2015: 975075.
 
13.
Dhital A, Spalton DJ, Goyal S, et al.: Calcification in hedrophilic intraocular lenses associated with injection of intraocular gas. Am J Ophthalmol. 2012; 153: 1154–1160.
 
14.
Khan MI, Muhtaseb M: Opacification of the intraocular lens implant following uneventful Descemet’s stripping endothelial keratoplasty. Cont Lens Anterior Eye. 2011; 34: 92–93.
 
15.
Ahad A, Darcy K, Cook SC, et al.: Intraocular lens opacification after Descemet stripping automated endothelial keratoplasty. Cornea. 2014; 33: 1307–1311.
 
16.
Tuberville AW, Wood TO, McLaughlin BJ: Cytochrome oxidase activity of Fuchs’ endothelial dystrophy. Curr Eye Res. 1986 Dec; 5(12): 939–947.
 
17.
Hesham N, Schultze RL: Impact of Donor Characteristics on 2-Year Descemet Stripping Automated Endothelial Keratoplasty Outcomes in Patients With Fuchs Endothelial Dystrophy. Cornea. 2015; 34: 6–10.
 
18.
Gartaganis SP, Kanellopoulou DG, Mela EK, et al.: Opacification of hydrophilic acrylic intraocular lens attributable to calcification: investigation on mechanism. Am J Ophthalmol. 2008; 146(3): 395–403.
 
19.
Gartaganis SP, Prahs P, Lazari ED, et al.: Calcification of Hydrophilic Acrylic Intraocular Lenses With a Hydrophobic Surface: Laboratory Analysis of 6 Cases. Am J Ophthalmol. 2016; 168: 68–77.
 
20.
Verdaguer P, Gris O, Casaroli-Marano RP, et al.: Intraocular Lens Opacification after Endothelial Keratoplasty as Analyzed by Environmental Scanning Electron Microscopy. Cornea. 2015; 34: 972–975.
 
21.
Werner L, Wilbanks G, Nieuwendaal CP, et al.: Localized opacification of hydrophilic acrylic intraocular lenses after procedures using intracameral injection of air or gas. J Cataract Refract Surg. 2015; 41(1): 199–207.
 
22.
Morgan-Warren PJ, Andretta W, Patel AK: Opacification of hydrophilic intraocular lenses after Descemet stripping automated endothelial keratoplasty. Clinical Ophthalmology. 2015; 9: 277–283.
 
23.
Neuhann IM, Neuhann TN, Rohrbach JM: Intraocular Lens Calcification After Keratoplasty. Cornea. 2013; 32: e6–e10.
 
24.
Nieuwendaal CP, van der Meulen IJE, Patryn EK, et al.: Opacification of the Intraocular Lens After Descemet Stripping Endothelial Keratoplasty. Cornea. 2015; 34(11): 1375–1377.
 
25.
Schmidinger G, Pemp B, Werner L: Opacification of an intraocular lens: calcification of hydrophilic intraocular lenses after gas tamponade of the anterior chamber. Ophthalmologe. 2013; 110(11): 1066–1068.
 
26.
Habib NE, Freegard TJ, Gock G, et al.: Late surface opacification of Hydroview intraocular lenses. Eye (Lond) 2002; 16(1): 69–74.
 
27.
Bompastor-Ramos P, Pavoa J, Lobo C, et al.: Late postoperative opacification of a hydrophilic-hydrophobic acrylic intraocular lens. J Catarct Refract Surg. 2016; 42: 1324–1331.
 
28.
Green FG, Werner L, Apple DJ, et al.: An issue resolved. The Hydroview intraocular lens: development, early reports of calcification and subsequent actions. White paper Bausch and Lomb; 29 July, 2003.
 
29.
Fellman MA, Werner L, Liu ET, et al.: Calcification of a hydrophilic acrylic intraocular lens after Descemet-stripping endothelial keratoplasty: Case report and laboratory analyses. J Cataract Refract Surg. 2013; 39: 799– –803.
 
30.
Guan X, Tang R, Nancollas GH: The potential calcification of actacalcium phosphate on intraocular lens surfaces. J Biomed Mater Res. 2004; 71A: 488–496.
 
31.
Werner L, Hunter B, Stevens S, et al.: Role of Silicon Contamination on Calcification of Hydrophilic Acrylic Intraocular Lenses. Am J Ophthalmol. 2006; 141: 35–43.
 
32.
Ohrstrom A, Svensson B, Tagenfeldt S, et al.: Silicone oil content in ophthalmic viscosurgicsl devices. J Catarct Refract Surg. 2004; 30: 1278– –1280.
 
33.
Patryn E, van der Meulen IJE, Lapid-Gortzak R, et al.: Intraocular Lens Opacifications in Descemet Stripping Endothelial Keratoplasty Patients. Cornea. 2012; 31: 1189–1192.
 
34.
Tandogan T, Khoramnia R, Choi CY, et al.: Optical and material analysis of opacified hydrophilic intraocular lenses after explantation: laboratory study. BMC Ophthalmology. 2015; 15: 170–177.
 
35.
Dagres E1, Khan MA, Kyle GM, et al.: Perioperative complications of intraocular lens exchange in patients with opacified Aqua-Sense lenses. J Cataract Refract Surg. 2004 Dec; 30(12): 2569–2573.
 
eISSN:1689-362X
ISSN:1505-2753
Journals System - logo
Scroll to top